Color Analyzer – AQUAS SMR47 Series

Color Analyzer – AQUAS SMR47 Series, Color Analyzer terhubung langsung melalui antarmuka komunikasi RS485, menyediakan data proses yang sederhana, andal, dan hemat biaya dengan kemampuan pemantauan, kalibrasi, konfigurasi, dan diagnostik jarak jauh. Perumahan Color Meter dalam selungkup bukti IP68 yang kuat, kabel yang diperkuat dengan kekuatan tarik 1500 N Kevlar, transmisi data digital hingga 1,2 km, pemancar ini ideal digunakan dalam industri air/air limbah.

*Catatan

Untuk mengetahui informasi terkait harga, silahkan hubungi kami sekarang juga!

Fitur:

  • All-in-One Compact Housing, Built-in Transmitter and Sensors
  • Robust IP68 Water Submersible Protection, Directly Installed in the Field, No Cabinet Required
  • Plug & Play, On-line Realtime Measurement
  • Ultra Low Power Consumption, Ideal for Outdoor Applications
  • 1500 N Tensile Strength Kevlar Reinforced Cable
  • Surge Protection for Power and RS485 Communication
  • RS485 Digital Communication, Minimize Cabling and Engineering Cost
  • Standard Modbus RTU Protocol, Direct Connected with PLC, HMI
  • Sapphire Glass Windows to Prevent Scratch
  • Innovative Nano Coating to Remain Window Clean
  • Auto Cleaning Wiper, Less Maintenance
  • Turbidity Measurement and Compensation
  • Onboard Memory Allowing Users Easily Calibrate and Configure Sensor at Lab and Distribute to Various Fields and Sites
  • AQCFG Software Tool for Data Monitoring, Calibration, Configuration and Diagnosis
  • IECEx/ATEX Ex ia IIB T5 Ga Explosive Proof Certification

Aplikasi

  • Air minum
  • Air permukaan
  • Industri
  • Pengolahan air
  • Air limbah
Bagikan ke:

Spesifikasi

MEASUREMENT METHOD
The analyzer consists of a light source, a sample cell, and a light detector (photo detector). Incident light is absorbed by the sample, and the light is
measured by the detector. The value depends on the amount of absorbance in the sample. The transmitter uses a long life LED and the 180° light
method in accordance with International standard to assure accurate measurement values. An automated mechanical wiper is utilized to remain
surface clean and remove air bubbles of the optical window in order to maximize the accuracy and minimize the maintenance requirement.
INSTALLATION
Submersible, flow through, pipe Insertion
GENERAL
Output Signal Rs485 (Modbus RTU protocol), 19,200 bps, 8 data bits, no parity, 1 stop bit; 4~20 mA (optional)
Data Resolution 16 bits (0.001% FS)
Surge Protection 4000 V DC
Power 5~12 VDC, 45 mA
Protection Polarity, Overload, Short circuit
Safety CE, FCC
COLOR MEASUREMENT
Range SMR47-1: 0~100 Hazen (Pt/Co, APHA, PCU)(100 mm)
SMR47-2: 0~500 Hazen (Pt/Co, APHA, PCU)(20 mm)
SMR47-3: 0~2,000 Hazen (Pt/Co, APHA, PCU)(5 mm)
Accuracy ±5%±1 Hazen in standard solution
Resolution 0.01 Hazen (Pt/Co)
Repeatability ±3% Hazen in standard solution
Light source LED 390 nm
Beam angle 180°
Light Path Length 5, 20, 100 mm
Process Flow Rate Max. 3m/sec
Operation pressure Max. 10 Kgf/cm²
Operating temperature 0 ~ 60°C
Response time 1 sec
TURBIDITY MEASUREMENT
Light source LED 880 nm
Range 0~1,000 NTU
Accuracy ±3% measured value
Resolution 0.001 NTU
Repeatability ±1% measured value
PROTECTION IP68
CONNECTION 1”-11 PT, 5-pin M16 plug
HOUSING SS316L; Titanium
DIMENSIONS Φ 50 X 305 mm
WEIGHT analyzer: approx. 1 Kg (SS316L); approx. 0.8 Kg (Titanium)
Cable: 80 g/m